Product IntroductionThe shielding design should fully consider the direction and range of exposure to the useful radiation beam, the operational load of the device, and outdoor conditions. The walls in the direction of the useful beam projection should be designed according to the primary shielding requirements, while those in other directions should be designed according to the requirements for leakage and scattered radiation shielding.
Product Performance- The door surface can be treated with sandblasting and fluorocarbon coating, and the motor shall meet the IP54 dustproof and waterproof rating.
- Door Integrity: The shielding materials shall be continuous and intact. The lead sheet inside the door body shall be tightly fitted to the frame to prevent radiation leakage caused by looseness or gaps.
- Process: 30mm Pb + 120mm boron-containing polyethylene paraffin.
- Material: Door panel: 1.0 mm thick; Door body: 10 mm Q235 steel plate.
- Motor: Adopts a brand-name motor, equipped with high-load-bearing rails and cover plates.
- Protection: Equipped with a special anti-fingerprint mirror finish belt for radiation protection.
- Color: 304 anti-fingerprint stainless steel or AkzoNobel electrostatic spraying sheet.
- Frame Requirement: The frame is supported by No.24 channel steel.
Technical ParametersTechnical parameters are presented in the product documentation/figures on the page (illustrative technical image included on the original page).
Application Fields- Department of Radiology:Comprehensive department integrating imaging diagnosis and interventional therapy. Radiological diagnosis employs X-rays, gamma rays and other radiation to penetrate the human body, rendering images of internal structures and organs for evaluation of anatomy, function and pathology.
- Department of Radiation Oncology:Localized therapeutic modality using ionizing radiation for cancer treatment; sources include radioactive isotopes and particle beams generated by therapy units and medical accelerators.
- Department of Nuclear Medicine:Adopts nuclear science technologies for disease diagnosis and treatment; integrates imaging, therapeutic services, research and teaching.
Caractéristiques / spécifications techniques- Shielding composition: 30 mm lead (Pb) + 120 mm boron-containing polyethylene paraffin.
- Door panel thickness: 1.0 mm.
- Door body material: 10 mm Q235 steel plate.
- Surface treatment options: sandblasting + fluorocarbon coating; 304 anti-fingerprint stainless steel or AkzoNobel electrostatic spray finish.
- Motor rating: meets IP54 dustproof and waterproof standard; uses brand-name motor with high-load-bearing rails and cover plates.
- Integrity requirement: continuous shielding materials; lead sheet tightly fitted to frame to avoid gaps and leakage.
- Protection features: special anti-fingerprint mirror finish belt for radiation protection.
- Frame support: No.24 channel steel.